Notice:

Cooling water circuit

Risk of oil leaking from the refrigeration system (compressor) of the
recirculating cooler into the cooling water in case of a fault in the cooling water
circuit!
Observe the laws and regulations of the water distribution company valid in
the location where the unit is operated.


Notice:

Danger of corrosion of heat exchanger due to unsuitable quality of
cooling water.
• Due to its high content of lime, hard water is not suitable for cooling and

causes scale in the heat exchanger.

• Ferrous water or water containing ferrous particles will cause formation of

rust even in heat exchangers made of stainless steel.

• Chlorinated water will cause pitting corrosion in heat exchangers made of

stainless steel.

• Due to their corrosive characteristics, distilled water and deionized water

are unsuitable and will cause corrosion of the bath.

• Due to its corrosive characteristics, sea water is not suitable.

• Due to its microbiological (bacterial) components, which settle in the heat

exchanger, untreated and unpurified river water and water from cooling
towers is unsuitable.

• Avoid particulate matter in cooling water.

• Avoid putrid water.

Recommended quality of cooling water:

pH

7.5 to 9.0

Sulfate [SO4 2- ]

< 100 ppm

Hydrocarbonate [HCO 3-]/sulfate [SO4 2-]

> 1 ppm

Hardness [Ca 2+, Mg 2+]/[HCO 3-]

> 0.5

°dH

Alkalinity

60 ppm < [HCO 3-] < 300 ppm

Conductivity

<

500 μS/cm

Chloride (Cl -)

< 50 ppm

Phosphate (PO4 3-)

< 2 ppm

Ammonia (NH3)

< 0.5 ppm

Free chlorine

< 0.5 ppm

Trivalent iron ions (Fe 3+)

< 0.5 ppm

Manganese ions (Mn 2+)

< 0.05 ppm

Carbon dioxide (CO2)

< 10 ppm

Hydrogen sulfide (H2S)

< 50 ppm

Content of oxygen

< 0.1 ppm

Algae growth

impermissible

Suspended solids

impermissible