JULABO F38-EH Refrigerated and Heating Circulators User Manual
Page 9

EH
9
2.4. Technical
specifications
F38-EH
Working temperature range
°C
-35 ... 80
Temperature stability
°C
±0.03
Temperature selection
digital
Temperature indication
LED
Resolution °C
0.1
ATC - Absolute Temperature Calibration
1-point
Temperature control
PID1
Heater wattage (at 230 V)
kW
2.0
Cooling capacity
°C
Medium ethanol
kW
+20 0 -20
0.92 0.66 0.32
Refrigerant R404A
Circulating pump:
discharge, max.at 0 bar
l/min
15
pressure, max. at 0 l
bar
0.35
Electrical connections:
External alarm device
Vdc/mA
24-0 / max. 25
Overall dimensions (WxDxH)
cm
44x64x86
Bath opening (WxL)
cm
35x41
Bath depth
cm
27
Filling volume from ... to
liters
38 ... 45
Weight kg
72
Ambient temperature
°C
5 ... 40
Mains power connection 230 V/50 Hz
V/ Hz
207-253 / 50
Current input (at 230 V)
A
13
Mains power connection 230 V/60 Hz
V/ Hz
207-253 / 60
Current input (at 230 V)
A
13
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.