Setting of temperatures – JULABO FP89-ME Ultra-Low Refrigerated Circulators User Manual

Page 24

Advertising
background image

Setting of temperatures

24

Control of the FP89 cooling machine:

The mains switch is turned on. After pressing the start/stop button, the
cooling machine is automatically switched on or off by the circulator. In
the start mode of the circulator, the cooling machine runs continuously.
Immediately after starting the cooling machine, the output of the system
is reduced to a minimum value for a defined period in order to equilibrate
the cooling system. Afterwards the output of the cooling machine is
ramped up to the capacity required by the circulator.

A control of the capacity limitation subject to the temperature of the bath
is stored in the circulator. This is used to increase the operational
reliability and does not affect the maximum capacity of the cooling
machine.















7.

Setting of temperatures

7.1.

Using the pre-settings in the

menu

Press the

key to call up the menu for temperature selection.

3 different working temperatures can be adjusted. Their values are freely
selectable within the operating temperature range.

The temperatures can be set in start or stop mode.

Press

key if a value is to be retained

Setting of working temperature in the

menu

Werkseinstellungen:
SETPNT 1

25 °C

SETPNT 2

37 °C

SETPNT 3

70 °C

1. Press the key

. The value flashes

2. Select SETPOINT 1 or 2 or 3 using the key

or

.

3. Confirm by pressing the

key.

The circulator uses the new working temperature value for temperature
control.

Advertising