JULABO MB Heating Immersion Circulator User Manual
Page 16

Preparations
16
5. Preparations
5.1. Installation
Caution:
Securely fix the immersion circulator. The heater may not be in contact with the
wall of the bath tank. Keep a distance of at least 15 mm.
Units not adequately fixed may drop into the bath tank.
D a n g e r o f e l e c t r i c s h o c k !
First pull out the power plug to disconnect the unit from the power supply net.
Then take the immersion circulator out of the bath tank.
Make a service technician check the instrument before it is used again.
Place the unit on an even surface on a pad made of non-
flammable material.
The heating immersion circulator is mounted using a bath
attachment clamp (21) designed for bath wall thicknesses
up to 26 mm.
21
22
Use the two sleeves (22) supplied with the unit to reduce
the immersion depth from 165 mm to 145 mm (see
drawing).
For use with glass vessels an upright stand rod,
available as optional accessory (order no. 8 970 022)
may be attached.