Operating procedures – JULABO LC 6 Programmable Controller User Manual
Page 15

15
5. Operating
procedures
5.1. Installation
The unit should be set up at a dry location.
Place the unit in an upright position and do not obstruct the
ventilation.
A wall distance of at least 10 cm must be maintained for
ventilation, allowing internal heat
to be conducted away from the unit.
If one or more temperature controllers are set up in a
cabinet for example, take care of good ventilation (waste
heat per unit = approx. 60 Watts).
The ambient temperature must not exceed 35 °C. Ambient
temperatures above 35 °C result in a failure of the unit.
Do not set up the unit in the immediate vicinity of heat
sources and do not expose to sun light.
5.2.
Connecting a heating device
Caution:
Set up the heating device according to the instructions or securely fix the unit in the
bath tank using appropriate means. Danger of burning and fire!
Connect the power plug to the grounded mains socket (36) on
the rear of the programmable controller.
Caution:
Max. resistive load 3000 W at 230 V. Max. current 16 A.
Max. resistive load 1400 W at 115 V. Max. current 12 A.
+