Kds ezflow 2040, Operation settings and procedures – KD Scientific EZFLOW 2040 User Manual

Page 24

Advertising
background image

22

KDS EZFlow 2040

Operation Settings and Procedures

(Continued)

9. OCCLU. LEVEL (OCCLUSION LEVEL)

MENU

ENTER

SET

Screen 21

Screen 22

HIGH: If the pump is not sensitive enough to trigger occlusion alarm with certain I.V sets,

HIGH occlusion sensitivity level should be selected.

MEDIUM: MEDIUM occlusion sensitivity level is the factory default setting. This occlusion

sensitivity level is preferred for most scenarios.

LOW: If the pump is too sensitive enough to trigger occlusion alarm, LOW occlusion sensitivi-

ty level shall be selected.

LOWEST: If the pump is still too sensitive and frequently sets off occlusion alarm with other

I.V sets, LOWEST occlusion sensitivity level shall be selected.

WARNING: The occlusion alarm will not be triggered under LOWEST OCCLUSION SEN-

SITIVITY LEVEL for most pump use IV sets.

1)

Press

to enter MODE SETTING

status.

2)

Press

to item No.8 OCCLU. LEVEL

(OCCLUSION LEVEL).

3)

Press

to enter ADJ. OCCLUSION

SENSITIVITY menu (screen 21 and

screen 22). There are 4 levels of

occlusion sensitivity to adjust: