TROUBLE SHOOTING

26

Z306 Instruction

manual.docx

© Hermle Labortechnik GmbH

4. Trouble Shooting

4.1 Error messages: Cause / Solution

The error messages are listed to help localize possible errors faster.
The diagnose referred to this chapter may not always be the case, as they are only theoretically
occurring errors and solutions.
Always, please keep us informed about any kind of error occurring, which is not listed in this chapter.
Only through your information we are able to improve and complete this operation manual.

Many thanks in advance for your support.

HERMLE Labortechnik GmbH


4.2 Survey of possible error messages and their solutions

4.2.1 Lid release during power failure (Emergency Lid Release)

In case of power failure or malfunction, the lid of the centrifuge can be opened manually in order to
protect your samples.

Please proceed as follows (see Figure 26):

Switch the centrifuge off and unplug the power cord, wait until the rotor stands still (this

may take several minutes)

At the right side of the centrifuge there is a plastic stopper (Figure 26). Remove this stopper,
which is connected to the lid lock, horizontally from the housing until the centrifuge lid opens.

• Now open the lid of the centrifuge.