Labnet Z400K High Capacity Centrifuge User Manual

Page 22

Advertising
background image

22 E

Z 400 K / March 1997

Error no.: 70

Error no.: 70

Error no.: 70

Error no.: 70

Error no.: 70

Interface of the frequency converter

Interface of the frequency converter

Interface of the frequency converter

Interface of the frequency converter

Interface of the frequency converter

Reason:

Communication of controller, power board, interface cable

and frequency converter is not working

Action:

Call service

Error no.: 82-83

Error no.: 82-83

Error no.: 82-83

Error no.: 82-83

Error no.: 82-83 Cutoff power board - frequency converter

Cutoff power board - frequency converter

Cutoff power board - frequency converter

Cutoff power board - frequency converter

Cutoff power board - frequency converter

Reason:

Overcurrent or undervoltage due to power supply fluctuations

Action:

Restart the unit, take care that the power supply is stable

Error no.: 84

Error no.: 84

Error no.: 84

Error no.: 84

Error no.: 84

Overtemperature at the driving

Overtemperature at the driving

Overtemperature at the driving

Overtemperature at the driving

Overtemperature at the driving

Reason:

Temperature at the converter or motor too high

Action:

Switch off the centrifuge. Wait for about 15 min. and switch

the unit on again

Error no.: 85-87

Error no.: 85-87

Error no.: 85-87

Error no.: 85-87

Error no.: 85-87 Failures

Failures

Failures

Failures

Failures

Reason:

Internal defect

Action:

Call service

Error no.: 90

Error no.: 90

Error no.: 90

Error no.: 90

Error no.: 90

Emergency lid release

Emergency lid release

Emergency lid release

Emergency lid release

Emergency lid release

Reason:

The centrifuge lid has been opened by the emergency lid release

during the run

Action:

Close centrifuge lid. Danger of accident!

Reason:

Control switch of the lid lock is defective

Action:

Call service

Error no.: 94

Error no.: 94

Error no.: 94

Error no.: 94

Error no.: 94

Voltage loss during run

Voltage loss during run

Voltage loss during run

Voltage loss during run

Voltage loss during run

Reason:

The power supply is below the tolerance for a short moment

Action:

Wait till stillstand of the rotor. Open centrifuge lid after the yellow

LED "lid" is lighting. Switch off and on the main switch.