USING YOUR MICROSENTRY™ REFRIGERATOR CONTROL

See the “care and cleaning” section for cleaning instructions.

Alarm Mute

Press any key to mute the audible portion of an alarm.

NOTE-This action will only mute the alarm. If the condition that

caused the alarm continues, the alarm code will continue to flash

and will sound for 20 seconds every 60 minutes.

Turning Refrigerator Off

To turn refrigerator off, press and hold “ON/OFF” button for

three (3) seconds. “OFF” will appear on the display.

Additional MicroSentry™ Features

Refer to Appendix A for details on additional features available.

Orange Wire / NC

Blue Wire / COM

Yellow Wire / NO

Remote Alarm Connections

Maximum Contact Rating: 125-250V 5A

Remote Alarm Relay

Remote Alarm Interface

The MicroSentry™ control provides the ability to monitor alarm

conditions via a set of dry relay contacts. Alarm functions that

may be monitored by the dry contacts are: door ajar, high / low

temperature, senor fault, and power failure. Note: Intentionally

turning the unit off will disable all alarm outputs on the remote in-

terface except power failure. It is recommended that the remote

monitoring device(s) have an uninterruptible power source.

NOTE

The alarm interface follows the state of the alarms. Muting the

alarm will not change the state of the relay contacts.

Figure 8