2.10 Ambient conditions

►

The appliance may only be used in enclosed areas and under the following ambient condi-

tions:

Ambient temperature

15 °C to 28 °C (to 34 °C with limited temperature

and humidity range)

Humidity rh

max. 70 %, non-condensing

Overvoltage category

II

Pollution degree

2

Altitude of installation

max. 2,000 m above sea level

►

The appliance may not be used in areas where there is a risk of explosion. The ambient air

must not contain any explosive dusts, gases, vapours or gas-air mixtures. The appliance is

not explosion-proof.

►

Heavy dust production or aggressive vapours in the vicinity of the appliance could lead to

sedimentation in the interior and, as a consequence, could result in short circuits or dam-

age to electrical parts. For this reason, sufficient measures to prevent large clouds of dust

or aggressive vapours from developing should be taken.

2.11 Scope of delivery

►

Power cable

►

Sliding grid (load capacity 30 kg each)

►

Water tank with connection hose

►

USB storage medium with software and AtmoCONTROL manual

►

The operating instructions at hand

►

Calibration certificate

►

Separately packaged fastening material for wall mounting (see page 20)

►

Tank holder (only for appliances of size 750, see page 22)

2.12 Optional accessories

►

USB to Ethernet converter (Fig. 8). Makes it

possible to connect the Ethernet connection

interface (see page 12) to the USB port of

a computer / laptop.

►

Reinforced, sliding steel grids with a load

capacity of 60 kg each (for appliance size

110 and larger)

Fig. 8

Converter USB to Ethernet