4.5 Troubleshooting

Is the unit connected to the mains supply? If not, connect unit to the mains.

Is the unit switched on? If not, switch unit on.

Are all tubes correctly connected? If not, connect tubes.

Is the waterbath heater switched on? If not, switch waterbath heater on.

It is possible that the tubes in the tube suspender are blocked. In this case, replace the

tube suspender.

If the error cannot be eliminated, the cause is probably a defect in the unit. In this case you

should contact customer service (see page 2).

4.6 Dismantling the cooling unit

1. Switch off the cooling unit; pull out the

mains plug.

2. Remove the tube suspender from the water-

bath. Keep the tubes in place when doing

this.

3. Attach the closing panel I1/J1 with two sup-

plied screws on the right side of the inside of

the lid (Fig. 22). (This is to prevent heat loss

through the cut-out in the lid for the tube

suspender.)

The waterbath can now be used again without

the cooling unit.

5. Disposal

Have the unit disposed of as scrap metal by a specialist disposal company.

I1/J1

I1/J1

Fig. 22

Assemble the closing panel