Cable specification – MTS FlexTest Models 200 User Manual

Page 100

Advertising
background image

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Model 494.44 Two-Station System Board

FlexTest Controller Configurations

100

UPS Connections for the Model 494.44 System Board (FT40)

The following drawing shows UPS connections for the Model 494.44 System
Board. Once connected, you must use your controller software to configure the
various UPS options.

Cable specification

To maintain EMC compliance, the UPS cable that connects to the load frame Y
cable must comply with the following specifications:

Connector type–9-pin, type D, male EMI connector.

Backshell–EMI metallized plastic or metal.

Cable–26–22 AWG, four-conductor with overall braided shield, with the braided
shield connected to the metallized backshell at the chassis.

Controller Interlock
Open = Controller Interlock

494.44 System

I/O Board

Fused 24 V DC
to Crosshead

J29A

L Frame

1

4

3

2

7

5

8

13

12

14
11
15

9
6

10

Crosshead Interlock
Common

+24V

External Program
Interlock Source
Open=Program Interlock

To J25 HPU

J25

HPU

6

5

HPU E Stop

+24V

HSM High

Crosshead Unlock Power
Crosshead Unlock
Power Common

Load Frame

J29A

Reserved

Power

100-240 VAC

50-60 Hz, 1-2 A

DA Output

J24

1 2 3 4 5 6 7 8

1 2 3 4 5 6 7 8

LAN 2

LAN 1

DEBUG

MOT

OROLA

10/100 BASE T 10/100 BASE T

SCSI

BUSY

PIB

BUSY

PCI MEZZANINE CARD

PCI MEZZANINE CARD

J55 Dig Out

Dig In J54

J43A

J43B

J49

Aux Pwr

Estop/Run J23

Interlock

J29A

J29B

Load Frame

J28 HSM

A

-B

HPU J25

UPS AC Fail*

UPS

UPS
AC Fail

Active Alarm:
Open=Logic1

UPS Battery Low*

* UPS Inputs

are only available

on connector J29A

Load

Frame

UPS

To

494.44

J29A

Custom J29 Y Cable
(PN 100-197-651)
required for UPS systems.

D9S

2

7

8

4

D9P

2

7

8

4

D15S

UPS
Battery Low

D15P

UPS Cable

Load

Frame

Cable

D15P

Advertising