MTS FlexTest® Models 40/60/100/200 Controller Hardware

Model 494.40 I/O Carrier Board

VME Bus Boards

124

How to Install a Shunt-Calibration Resistor on an I/O Carrier Board

The Model 494.40 I/O Carrier board has a front-panel socket where you can
insert shunt-calibration plug assembly for use with DUC cards. Each socket is
associated with one of the eight RJ-50 connectors on the front panel.

1. Determine the RJ-50 connector(s) used by the transducer(s).

Note

Each mezzanine-card slot on the I/O carrier board connects to two RJ-50
connectors on the front of the I/O carrier board.

2. Solder the shunt-calibration resistors to an MTS shunt-calibration plug

assembly (MTS 11-433-826).

Note

The shunt-calibration plug assembly includes plug assemblies for eight
shunt resistors. Each resistor should be labeled with its resistance value
and transducer serial number.

3. Insert the shunt-calibration plug assemblies into the front-panel sockets.

Note

If you use MTS TEDS modules or MTS transducers with integrated
shunt-calibration resistors, you must insert a jumper plug (MTS 100-188-
097) into the socket for each transducer input where you will use the
integrated shunt-calibration resistor.

1

2

3

4

5

6

7

8

J1A

J1B

J2A

J2B

J3A

J3B

J4A

J4B

Connector

Numbering

Shunt-Calibration

Plug Assembly

2

1

3

1

2

3

4

5

6

7

8

J1A

J1B

J2A

J2B

J3A

J3B

J4A

J4B

Card Slot 1
uses RJ-50 connectors
J1A and J1B

Card Slot 2
uses RJ-50 connectors
J2A and J2B

Card Slot 3
uses RJ-50 connectors
J3A and J3B

Card Slot 4
uses RJ-50 connectors
J4A and J4B

Advertising