MTS FlexTest® Models 40/60/100/200 Controller Hardware

Mezzanine Cards

Mezzanine Cards

134

About Mezzanine Cards

You can install any of the mezzanine cards listed in the following table in
one of the four card slots on the Model 494.40 I/O Carrier board.

Damage can occur if you connect a high-level signal to a mezzanine
card when the chassis power is off.

Improper operation can damage mezzanine cards.

Do not apply a high-level signal to a mezzanine card when the chassis power
is off. Disconnect power from any externally powered devices before you
switch the chassis power off.

Connections

Each card slot has two front-panel RJ-50 connector on the I/O carrier board
where you can connect external devices or transition boards.

Setup

You can configure mezzanine cards using hardware switches on the card,
hardware-mapping software, and control software. Shunt-calibration and
bridge-completion resistors for DC conditioners are installed on the Model
494.40 I/O Carrier board.

CAUTION

1

2

3

4

5

6

7

8

J1A

J1B

J2A

J2B

J3A

J3B

J4A

J4B

Model 494.40

I/O Carrier Board

Mezzanine Card Slot 1
uses RJ-50 connectors
J1A and J1B

Mezzanine Card Slot 2
uses RJ-50 connectors
J2A and J2B

Mezzanine Card Slot 3
uses RJ-50 connectors
J3A and J3B

Mezzanine Card Slot 4
uses RJ-50 connectors
J4A and J4B

Note

Typically, conditioner
cards are installed in
slots 1 and 2,
speciality cards (such
as D to A or
transducer
conditioners) are
installed in slot 3, and
valve drivers are
installed in slot 4.