Shunt calibration, Full-bridge configuration – MTS FlexTest Models 200 User Manual

Page 148

Advertising
background image

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Digital Universal Conditioner Mezzanine Cards

Mezzanine Cards

148

Digital Universal Conditioner (DUC) Bridge Connections

The following figures show the full-, half-, and quarter-bridge configurations
for a DUC that is configured as a DC conditioner.

Shunt calibration

The shunt-calibration resistor (R Shunt) sockets are located on the front of
the I/O carrier board. Use system hardware-mapping software to specify
which leg of the bridge is shunted during calibration.

Note

If you use MTS TEDS modules or MTS transducers with integrated
shunt-calibration resistors, you must insert a jumper plug (MTS part
number 100-188-097) into the socket for each transducer input
where you will use the integrated shunt-calibration resistor.

Full-bridge

configuration

The following figure shows a full-bridge configuration.

-

Ex

-Ex Sense

I/O Carrier Board

Front-Panel

RJ-50 Connector

Transducer

RJ-50

Cable

Plug

2

1

Cable

Shield

3

Cable Grounding

1 The cable shield connects to the metal

shielding on the RJ-50 cable plug.

2 The cable plug shielding connects to the

I/O carrier board body.

3 The I/O carrier board connects to earth

ground through the chassis.

DUC

A to D

SW1B

3

8

6

4

7

5

9

2

+FBR

R

Shunt

+Ex Sense

+Excitation

-Excitation

-Ex Sense

+FB

-FB

-FBR

IEEE

1451.4

Circuit

+ IEEE 1451.4 Class 2

10

1

- IEEE 1451.4 Class 2