Model 494.25 Single DUC Card

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Mezzanine Cards

155

Model 494.25 Single DUC Card Settings

Hardware

configuration

The conditioner on this card has a two-position switch that is used to set up
the following DC conditioner functions:

Switch A connects a bridge completion resistor to -FBR to complete a
quarter-bridge circuit.

Switch B provides a ground reference for an external transducer by
connecting -FBR to ground.

Software configuration

After you set up and install the hardware, you must configure your control
software settings for the conditioner.

Note

For detailed information on software configuration, calibration, and
tuning, see the control software and tuning/calibration
documentation.

1 2

Switch Settings

A = Bridge Completion Switch
The ON position connects a
user-installed bridge completion
resistor to complete a quarter-
bridge circuit.

B = Grounding Switch
The ON position connects -FBR
line to ground. (See the bridge
completion drawings for details.)