MTS FlexTest® Models 40/60/100/200 Controller Hardware

Model 494.45 8-Input A/D Converter Card

Mezzanine Cards

178

Model 494.45 8-Input A/D Card Connections

Analog input signals are typically routed to the A/D converter card through a
Model 494.75 8-Input BNC Transition Board. The two transition board
outputs (J11, J12) connect to the I/O carrier board RJ-50 connectors used
with the A/D Converter card slot (J3A and J3B).

The front-panel sockets on the I/O carrier board only accept cabling
with 10-pin, shielded, RJ-50 connectors.

The use of other RJ connector types (less than 10 pins or unshielded)
with the I/O carrier board can cause component damage.

Only use cables equipped with 10-pin, braided shielded, RJ-50 connectors
(with gray boot) with the I/O carrier board.

CAUTION

First

494.76

ANALOG

OUTPUT

CH 1 - 8

Last

J11

J12

1

2

3

4

5

6

7

8

J1A

J1B

J2A

J2B

J3A

J3B

J4A

J4B

Model 494.75 8-Input

BNC Transition Board

Model 494.40 I/O Carrier Board

Transition

Card Cage

VME Card Cage

Analog

Input

Devices

Input 1
Input 2
Input 3
Input 4
Input 5
Input 6
Input 7
Input 8

Input 1-4

Input 5-8

494.45 8-Input

A/D

Mezzanine Card