Specifications–model 493.72 digital i/o board – MTS FlexTest Models 200 User Manual

Page 210

Advertising
background image

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Model 493.72 Digital I/O Board

Digital I/O and Transition Boards

210

Specifications–Model 493.72 Digital I/O Board

Low current

Low-current applications that use the Model 493.31 DI/O Breakout Box
typically use the 12 VDC power from the Model 493.72 board.

High current

High-current applications that use the Model 494.31 DI/O Breakout Box
typically use the output circuits from the Model 493.72 board to drive the
high-current switching circuits on the breakout box.

For more information

For device wiring diagrams see:

“Model 493.31 16-Channel Low-Current DI/O Breakout Box”

on page 212

“Model 494.31 16-Channel High-Current DI/O Breakout Box”

on page 216.

Model 493.72 Digital I/O Board Specifications

P

ARAMETER

S

PECIFICATION

Digital Inputs

Connector J3 (D-37S)

Provides 16 optically isolated digital inputs:

Input ON Voltage = 2.7–26 V DC
(at 0.5 mA minimum)

Input OFF Voltage = <0.8 V DC

Input Resistance = 2 Kohm

Jumper Selectable Debounce Times:
20 ms, 10 ms, 1 ms, 0.1 ms

Digital Outputs

Connector J4 (D-37S)

Provides 16 optically isolated digital outputs:

Output Voltage = 5–30 V DC

Output Current = 20 mA maximum at 12 V DC

Advertising