MTS FlexTest Models 200 User Manual

Page 211

Advertising
background image

Model 493.72 Digital I/O Board

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Digital I/O and Transition Boards

211

How to Change the Debounce Settings on the Model 493.72 Digital I/O Board

If necessary, install jumpers on the Model 493.72 Digital I/O board to
change the debounce time for each group of four digital inputs.

Note

The board ships with no jumpers installed, which provides a default
debounce time of 20 ms for each input.

1. Locate the four jumper sockets (X2-X5) on the Model 493.72 Digital

I/O board.

Each jumper socket sets the debounce time for a group of four inputs.

2. Install jumpers to change the debounce time.

Model 493.72

Digital I/O Board

X2

X3

X4

X5

Inputs

1-4

Inputs

5-8

Inputs

9-12

Inputs

13-16

Jumper Sockets

Debounce Jumper Settings

20 ms

(default)

10 ms

1 ms

0.1ms

1

2

4

3