Model 493.74 two-station hsm interface board, Model 493.74 two-station hsm interface board 246 – MTS FlexTest Models 200 User Manual

Page 246

Advertising
background image

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Model 493.74 Two-Station HSM Interface Board

Digital I/O and Transition Boards

246

Model 493.74 Two-Station HSM Interface Board

About the Model 493.74 Two-Station HSM Interface Board

The Model 493.74 Two-Station HSM Interface
board is a double-width board that plugs into the
transition card cage on a Series 494 Chassis.
Each Model 493.74 HSM transition board
controls up to two stations.

Each station has the following connections:

24-volt hydraulic service manifold (HSM)
control (off/low/high or proportional) (J28
HSM)

E-Stop and crosshead controls for a load
unit control module (J29 Load Frame)

Inputs that can initiate station and program
interlocks (J43 Interlock)

Remote run/stop outputs for external
devices (J44 Run/Stop)

Auxiliary power outputs for external
devices (J49 Aux Power)

Intlk

J28B HSM

R

J43-1

J43-2

Run/Stop

J44-1

J44-2

Aux Power

J49-1

J49-2

Load Frame

J29-1

J29-2

J28A HSM

493.74

HSM

Station

J49 Aux Power
Fused auxiliary DC supply
voltages for external devices.
+ 5 V DC, +15 V DC,
-15 V DC, and +24 V DC

J28A/B HSM
24-volt control for HSM off/
low/high or proportional
solenoids.

J43-1/2 Intlk
Station and Program
Interlock inputs

J44-1/2 Run/Stop
Run/Stop output contacts
Interlock out (opto isolators)

J29-1/2 Load Frame
E-Stop and crosshead
controls for each station