Connections for the model 493.07 converter box – MTS FlexTest Models 200 User Manual

Page 278

Advertising
background image

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Model 493.07 HPU Converter Box

Digital I/O and Transition Boards

278

Connections for the Model 493.07 Converter Box

This section describes how to connect a Series 494 Chassis to a variety of
MTS hydraulic configurations.

Cable specifications

Use the following connectors and cables to connect the Series 494 Chassis to
your HPU with a Model 493.07 Converter Box:

Series 494 Chassis to

Model 493.07

Converter Box

J25 Connector–15-contact, type D, female EMI connector at J25 of the
Model 493.73 HPU Transition board in the rear panel of the Series 494
Chassis, or J25 on the Model 494.41 System board in the Model 494.04
chassis.

Backshell–EMI metal.

J1 Connector–14-contact, type CPC, male connector at J1 of the Model
493.07 Converter Box chassis.

Cable–24 AWG, 10-conductor with overall foil shield, (Carol C0745 or
equivalent) with drain wire connected to metallized plastic backshell at the
493.10 chassis, and pin 14 at the Model 493.07 Controller Box chassis.

Model 493.07 Converter

Box to HPU

J25 Connector–24-contact, type CPC, female connector at J25 of the Model
493.07 Converter Box chassis.

J1 Connector–14-contact, type MS connector at J1 of the HPU.

Cable–18 AWG, 14-conductor with overall foil shield, with drain wire
connected to pin 4 of connector J25 and pin A of the pump connector.

Direct HPU

connections

The following diagrams show how to connect the chassis to a hydraulic
power unit.

Single Series 494 Chassis with a Model 505 Pump or 24 V PLC pump (Model 506.52-.92)

J25

HPU

494

Chassis

HPU

039-713-7XX

Advertising