UPS System Requirements

MTS FlexTest® Models 40/60/100/200 Controller Hardware

FlexTest Controller Configurations

45

The addition of a low battery warning relay contact out from the UPS will
provide additional system safety protection by letting the controller know that the
UPS battery is low.

Specifications–UPS Systems Used with MTS Controllers

Any UPS used with an MTS system must comply with these specifications.

UPS power

requirements

The following UPS power requirements include power for the PC, MTS
Controller (one chassis), and 25% capability for other subsystems.

UPS Specifications

I

TEM

R

EQUIREMENT

Operating temperature

5–40 deg C

Operating humidity

5–85% non-condensing

Supported UPS output
voltage range

100–240 V AC nominal, single phase,
sine-wave output

Input/output frequency
range

50–60 Hz

Output load regulation

-/+5% nominal operating voltage (both in
battery and normal operation modes)

Switch-over time to battery
on power loss

0 s (recommended), < 6 ms (required)

UPS Power Requirements

C

ONTROLLER

T

YPE

C

ONTROLLER

C

HASSIS

UPS P

OWER

R

EQUIRED

FlexTest SE

493.02

1200 watt

FlexTest GT

493.10

2500 watt

Aero ST

493.20

3500 watt

Flextest 40

494.04

1200 watt

FlexTest 60

494.06

1800 watt

FlexTest 100

494.10

2500 watt

FlexTest 200

494.20

3500 watt