J29 Load Frame Connections for the Model 494.42 System Board

Connector J29 Load Frame provides an interface to connect one load frame.

Cable specification

To maintain EMC compliance, the J29 Load Frame cable must comply with the
following specifications:

Connector type–15-pin, type D, male EMI connector.

Backshell–EMI metallized plastic or metal.

Cable–18 AWG 8-conductor with foil shield, with the drain wire connected to a
metallized plastic backshell at the chassis.

HPU E Stop 2*

To HPU E Stop Chain

Controller Interlock
Open = Controller Interlock

494.42 System Board

Fused 24 V DC
to Crosshead

J29

1

4

3

2

7

5

8

13

11

15

12

14

Crosshead Interlock
Common

+24V

External Program
Interlock Source
Open=Program Interlock

To HPU E Stop Chain

HPU E Stop 1

+24V

HSM High

Crosshead Unlock Power
Crosshead Unlock
Power Common

Load Frame

J29

Pins 6,9,10 Reserved

Power

100-240 VAC

50-60 Hz, 1-2 A

DA Output

J24

1 2 3 4 5 6 7 8

1 2 3 4 5 6 7 8

LAN 2

LAN 1

DEBUG

MOT

O

ROLA

10/100 BASE T 10/100 BASE T

SCSI

BUSY

PIB

BUSY

PCI MEZZANINE CARD

PCI MEZZANINE CARD

J54

Dig In

J55 Dig Out

J43 INTLK

J56 UPS

J49

Aux Pwr

J23 Estop/Run

J29 Load Frame

J25 HPU

J28 HSM

* MTS Model 505 Pumps and older do not use the

HPU E-Stop 2 signal.