Jumper plug required, Cable specification, 42 system board – MTS FlexTest Models 200 User Manual

Page 78

Advertising
background image

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Model 494.42 Single-Station System Board

FlexTest Controller Configurations

78

Jumper plug required

If connector J29 is not used, you must install a jumper plug to maintain the
integrity of the interlocks. Use jumper plug part number 100-007-947 or jumper
pins: 3-4, 5-7, 8-13, and 11-15.

J43 Interlock Connections for the Model 494.42 System Board

Connector J43 Intlk provides one optically isolated interlock input and a relay-
contact interlock output per connector.

Cable specification

To maintain EMC compliance, the J43 Intlk cable must comply with the
following specifications:

Connector type–9-pin, type D, male EMI connector.

Backshell–EMI metallized plastic or metal.

Cable–shielded twisted pairs (24 AWG minimum), braided shield with shield
connected to the metallized backshell at the chassis.

Event Input

+24V

1
2

3

4

Interlock Disabled

+24V

1
2

3

4

Switch Contact

(Open = Interlock)

+24V

1
2

3

4

Logic Input

(0=Interlock)

Interlock In (Gnd)

Interlock Out (NO)
Interlock = Open

Interlock Out (NC)
Interlock = Closed

Common

- Interlock In

+ Interlock In

Interlock In (Pwr)

+24V

J43

Intlk

494.42 System Board

6

7

8

1

2

3

4

5

9

Interlock

Output

Interlock

Input

Interlock

J43

Software
Defined

Power

100-240 VAC

50-60 Hz, 1-2 A

DA Output

J24

1 2 3 4 5 6 7 8

1 2 3 4 5 6 7 8

LAN 2

LAN 1

DEBUG

MOT

OROLA

10/100 BASE T 10/100 BASE T

SCSI

BUSY

PIB

BUSY

PCI MEZZANINE CARD

PCI MEZZANINE CARD

J54

Dig In

J55 Dig Out

J43 INTLK

J56 UPS

J49

Aux Pwr

J23 Estop/Run

J29 Load Frame

J25 HPU

J28 HSM

Advertising