MTS FlexTest® Models 40/60/100/200 Controller Hardware

Model 494.44 Two-Station System Board

FlexTest Controller Configurations

90

J25 Hydraulic Power Unit Connections for the Model 494.44 System I/O Board

Connector J25 HPU provides 24-volt logic signals that control the hydraulic
power unit (HPU). The connector may be connected directly to MTS Series 505
HPUs and similar HPUs that use low-current, 24-volt input controls.

Note

Other MTS HPUs require the Model 493.07 HPU Converter Box to
convert the low-current HPU output signal to a signal that can drive the
HPU relay.

ON = Normal

ON = Normal

ON = HPU ON

Power

100-240 VAC

50-60 Hz, 1-2 A

DA Output

J24

1 2 3 4 5 6 7 8

1 2 3 4 5 6 7 8

LAN 2

LAN 1

DEBUG

MOT

OROLA

10/100 BASE T 10/100 BASE T

SCSI

BUSY

PIB

BUSY

PCI MEZZANINE CARD

PCI MEZZANINE CARD

J55 Dig Out

Dig In J54

J43A

J43B

J49

Aux Pwr

Estop/Run J23

Interlock

J29A

J29B

Load Frame

J28 HSM

A

-B

HPU J25

J25

HPU

Controller 24 Vdc

+24 V DC (from HPU)

HPU

ON

Sense

24 V DC (from HPU)

Start

Lo

High

HPU Over Temp

Open = Interlock Active

494.44 System Board

9

10

11

12

1

2

3

4

5

6

15

7

8

J25

HPU

HPU Low Level

Open = Interlock Active

Start

SSR

Low

SSR

High

SSR

24 V DC (from HPU)

HPU ON

Series 505 HPU or Equivalent

HPU E-Stop Out 1

* MTS Model 505 Pumps and older do not use the

HPU E-Stop 2 signal.

HPU E-Stop Out 2*

Open = HPU E Stop