J28 hsm – MTS FlexTest Models 200 User Manual

Page 92

Advertising
background image

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Model 494.44 Two-Station System Board

FlexTest Controller Configurations

92

J28 HSM Connections for the Model 494.44 System Board

HSM control (off/low/high or proportional) for both HSMs is software
configurable. The Model 494.44 board provides separate 24-volt, low-pressure
and high-pressure outputs that drive the HSM low- and high-pressure solenoids.
Proportional solenoid control provides a current output from 0 to 0.78 A.

Note

The Model 494.44 board can not be used with 115 V AC HSMs.
Applications that use 115 V AC HSMs require an external converter box
(such as a Model 413.08), which is used with this board.

Current
Amplifier

HSM B Proportional

Control

7

4

8

6

1

5

9

+24V

High

Low

Reserved

Reserved

HSM B (D9S)

Current
Amplifier

Reserved

Reserved

HSM A (D9S)

494.44 System Board

7

4

8

6

1

5

9

+24V

HSM A Low

Solenoid

HSM A High

Solenoid

High

Low

J28

(D15P)

6

5

7

2

1

3

4

14

13

15

10

9

11

12

HSM A Proportional

Control

HSM B Low

Solenoid

HSM B High

Solenoid

HSM

A

HSM

B

J28 HSM

Power

100-240 VAC

50-60 Hz, 1-2 A

DA Output

J24

1 2 3 4 5 6 7 8

1 2 3 4 5 6 7 8

LAN 2

LAN 1

DEBUG

MOT

OROLA

10/100 BASE T 10/100 BASE T

SCSI

BUSY

PIB

BUSY

PCI MEZZANINE CARD

PCI MEZZANINE CARD

J55 Dig Out

Dig In J54

J43A

J43B

J49

Aux Pwr

Estop/Run J23

Interlock

J29A

J29B

Load Frame

J28 HSM

A-B

HPU J25

Note: A custom Y Cable
and individual HSM
cables are required.

To

494.44

J28

Custom J28 Y Cable

(PN 100-185-473)

D9S

D9S

D15P

HSM A

HSM B