Rtd temperature accuracy, Model 525 operating instructions – Practical Instrument Electronics 525 Rev B User Manual

Page 6

Advertising
background image

Model 525 Operating Instructions

82 E. Main Street Suite 3.14 · Webster, NY 14580

Tel: 585-872-9350 · Fax: 585-872-2638

[email protected] · http://www.piecal.com/

Practical Instrument Electronics, Inc. Copyright ã 2011. All rights reserved.

525-9002 7/13/11

6-8

RTD Temperature Accuracy


The following charts give worst-case temperature accuracy based on stated resistance accuracy of ± (0.015 % + 0.05) Ω.

Pt100 TCR=0.00385

-200 to 850 °C

0.1

0.2

0.3

0.4

-200

0

200

400

600

800

Temperature °C

Er

ro

r °

C

Pt100 TCR=0.003902

-200 to 648.9 °C

0.1

0.2

0.3

0.4

-200

0

200

400

600

800

Temperature °C

Erro

r °

C

Pt100 TCR=0.003916

-200 to 648.9 °C

0.1

0.2

0.3

0.4

-200

0

200

400

600

800

Temperature °C

Erro

r °

C

Pt100 TCR=0.003926

-200 to 850 °C

0.1

0.2

0.3

0.4

-200

0

200

400

600

800

Temperature °C

Er

ro

r °

C

Cu10 TCR=0.00427

-200 to 260 °C

1.2

1.25

1.3

1.35

1.4

-200

-100

0

100

200

300

Temperature °C

Er

ro

r °

C

Cu50 TCR=0.00428

-50 to 150 °C

0

0.1

0.2

0.3

0.4

0.5

-50

0

50

100

150

Temperature °C

Erro

r °

C

Ni120 TCR=0.00672

-80 to 260 °C

0.07

0.08

0.09

0.1

-100 -50

0

50

100

150

200

250

300

Temperature °C

Er

ro

r °

C












Ni110 TCR=0.00530

-100 to 260 °C

0.08

0.1

0.12

0.14

0.16

-100

0

100

200

300

400

Temperature °C

Error °C

Advertising