Thermocouple ranges & accuracies – Practical Instrument Electronics 820 User Manual

Page 31

Advertising
background image

Page 29

Thermocouple Ranges & Accuracies

Based on ±(0.03% of 80 mV)

T/C

Degrees C

Range

°C

Degrees F

Range

°F

J

-200.0 to -150.0

±1.2°

-328.0 to -238.0

±2.0°

-150.0 to -50.0

±0.7°

-238.0 to -58.0

±1.3°

-50.0 to 100.0

±0.5°

-58.0 to 212.0

±0.9°

100.0 to 1200.0

±0.4°

212.0 to 2192.0

±0.8°

K

-230.0 to -150.0

±2.6°

-382.0 to -238.0

±4.7°

-150.0 to 0.0

±1.0°

-238.0 to 32.0

±1.8°

0.0 to 1100.0

±0.6°

32.0 to 2012.0

±1.1°

1100.0 to 1371.1

±0.7°

2012.0 to 2500.0

±1.2°

T

-260.0 to -230.0

±6.1°

-436.0 to -382.0

±11.0°

-230.0 to -150.0

±2.2°

-382.0 to -238.0

±4.0°

-150.0 to 50.0

±1.1°

-238.0 to 122.0

±2.0°

50.0 to 300.0

±0.5°

122.0 to 572.0

±1.0°

300.0 to 400.0

±0.4°

572.0 to 752.0

±0.7°

E

-240.0 to -150.0

±2.5°

-400.0 to -238.0

±4.5°

-150.0 to -50.0

±0.7°

-238.0 to -58.0

±1.1°

-50.0 to 150.0

±0.4°

-58.0 to 302.0

±0.8°

150.0 to 1000.0

±0.3°

302.0 to 1832.0

±0.6°

R

-18.3 to 50.0

±6.5°

-1.0 to 122.0

±11.7°

50.0 to 500.0

±3.7°

482.0 to 932.0

±6.6°

500.0 to 800.0

±2.2°

932.0 to 1472.0

±4.0°

800.0 to 1767.8

±2.0°

1472.0 to 3214.0

±3.5°

Advertising