Practical instrument electronics, Model 541 specifications – Practical Instrument Electronics 541 User Manual

Page 7

Advertising
background image

Practical Instrument Electronics

82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

Page 7

General
Accuracy

±0.005% of Range

Ranges

0.001 to 20.000 kHz
0.1 to 2000.0 Hz
0.01 to 200.00 Hz
0.1 to 2000 CPM (Counts-per-Minute)
1 to 20000 CPH (Counts-per-Hour

Operating Temperature Range

-25 to 60 °C (-10 to 140 °F)

Temperature Drift

± 0.01% of span outside of 23°C ±10 °C (73°C ±18 °F)

Relative Humidity Range

10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing

Size

7.00 X 3.30 X 2.21 inches (177.8 X 83.8 X 56.1 mm

Weight

12.0 oz (340 grams) (including battery)

Batteries & Battery Life

One “Alkaline 9V (6LR61), 45 Hours

Low Battery

Low battery indication with nominal 1 hour of life left

Protection against mis-connection

Over-voltage protection to 120 Vrms rated for 30 seconds

or 240 Vrms for 15 seconds

Display

High contrast graphic liquid crystal display.

Normal Mode/Common Mode Rejection

50/60 Hz, 50 dB/ 50/60 Hz, 90 dB

Read Frequency
Input Impedance

> 1 Megohms + 100pF

Signal Attenuation Adjustment Range

Adjustable trigger level with X1 (0.1 V peak to 12.0 V peak)

and X10 (1 V Peak to 120 V peak)

Note on reading slow sine waves

High signal noise and low slew rate (Volts-per-second) will

affect reading uncertainty

Source Frequency
Output Impedance

< 25 Ohms

Source Current

> 6 mA at 12 V peak-to-peak, 20 kHz

Amplitude Adjustment Range & Accuracy

0.1 to 12.0 V peak ±10% of setting

Square Wave (Pulse)
Rise/Fall Time
Frequency Jitter
Duty Cycle

Selectable Zero Based or Zero Crossing
<0.0001% of output V peak per second
<0.5 Least Significant Digit of full scale value
50% ±2%

Sine Wave
Offset and Zero Crossing Symmetry

<±10% of V peak output amplitude setting

Model 541 Specifications

(Unless otherwise indicated all specifications are rated from a nominal 23 °C,

70 % RH for 1 year from calibration)