Practical Instrument Electronics

82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

Page 4

SOURCE

Choose this function to provide a simulated thermocouple signal into controllers, temperature
transmitters, indicators or any input devices that measure thermocouple sensors..

1) Disconnect the thermocouple sensor from the

device to be calibrated.

2) Select “

SOURCE” with slide switch

w

.

3) Connect a thermocouple wire (matching the type

of wire to sensor being simulated) with miniature
male T/C connector to the inputs of the device
being calibrated, making sure to check polarity.
Millivolt outputs (without cold junction) may be
connected with a copper (white) miniature T/C
connector or the banana jacks with copper wire.

The output is adjusted in 0.1° (or 0.001 mV)
increments by turning the knob

e

while the

EZ-CHECK™ switch

q

is in the “HI”, “LO” or “SET”

position. Press and turn the knob for faster dialing
with 10° (or 0.100) increments.

The OVERLOAD indicator will light if excessive
voltage or current is detected by the calibrator.

Sourcing Thermocouple

AUTOMATED THERMOCOUPLE

CALIBRATOR

J‐T‐E‐K‐R‐S‐B‐N‐G‐C‐D‐L‐U‐P‐mV

READ

SOURCE

HI

SET

LO

MAX

READ

MIN

OVERLOAD

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Double Click

for Options

OFF

READ

HI

TYPE K °C

17.332mV 22.4°C

422.1

READ

Choose this function to measure temperatures with a thermocouple probe or sensor.

1) Disconnect the thermocouple sensor from any other device.
2) Select “

READ” with slide switch

w

.

3) Place the EZ-CHECK switch into the

READ position.

4) Connect a thermocouple probe (matching the type of wire to sensor being measured) with min-

iature male T/C connector to the sensor. Millivolt outputs (without cold junction) are connected
with a copper (white) miniature thermocouple connector or to the banana jacks with cooper wire.

The PIECAL 322 measures the temperature signal and constantly updates the display with the
current temperature reading. Move the EZ-CHECK switch to MAX to see the highest temperature
reading and to MIN to see the lowest temperature reading. Press and hold the knob

e

to clear the

MAX and MIN readings.

The OVERLOAD indicator will light if excessive voltage or current is detected by the calibrator.

Reading Thermocouple Sensors

Warranty

Our equipment is warranted against defective material and workmanship (excluding batteries) for a
period of three years from the date of shipment. Claims under warranty can be made by returning
the equipment prepaid to our factory. The equipment will be repaired, replaced or adjusted at our
option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our
warranty. No responsibility is accepted for damage, loss or other expense incurred through sale or
use of our equipment. Under no condition shall Practical Instrument Electronics, Inc. be liable for any
special, incidental or consequential damage.