RaySafe i2 System User Manual – Maintenance
Chrome parts should only be cleaned by rubbing down with a dry cloth. do not use abrasive polishes. to
preserve the finish, use non-abrasive wax.
dISInfeCtIon
those parts of the product that are suitable for such treatment, including accessories and connecting cables,
can be disinfected by wiping with a cloth dampened with a suitable agent. never use corrosive or solvent
disinfectants. If you are not sure about the properties of a disinfectant, do not use it.
WARNING
Do not use flammable or potentially explosive disinfecting sprays. Such sprays create vapors,
which can ignite, causing fatal or other serious personal injury.
CAUTION
Disinfecting a medical product room by means of sprays is not recommended, since the vapor
could penetrate the product, causing electrical short-circuits, metal corrosion or other damage
to the product.
If non-flammable, non-explosive spray disinfectants are to be used, the equipment must first be switched
off and allowed to cool. this prevents convection currents from drawing spray mist into the product. plastic
sheeting must be used to cover the product thoroughly, after which spraying can begin.
once all traces of the disinfectant vapor have dispersed, the plastic sheeting can be removed and the
equipment itself can be disinfected in the recommended way.
If a spray was used, you must be satisfied that all traces of the vapor have dispersed before switching the
product on again.