Setting up on a laboratory table

Fig.3

You can mount the SK 100 on your laboratory table. Fig. 3

*

Measure laboratory table thickness

*

Use screws with a maximum diameter of 8mm and a

length of the laboratory table thickness + 25 mm

*

Stand the SK 100 including collecting vessel on the

table

*

Mark the holes for the SK 100 for the table with ref-

erence to the holes on the motor


The distance between the holes on the front edge of the table

must be such that the collecting vessel or filter hose can be

mounted and dismounted easily.

Electrical connection

*

The voltage and frequency of the SK 100 can be found

on the nameplate.

*

Ensure that the values agree with the power supply avail-

able.

*

Connect the SK 100 to the power supply using the con-

necting cable supplied.

The power cable supplied does not have a plug because the

type of plug depends on the setting up location and the respec-

tive national regulation.

* When connecting the power cable to the power supply,

provide an external fuse in accordance with the regulations

of the setting-up location.

If the values are not observed, electronic and mechanical compo-
nents may be damaged.

Before first use, the direction of rotation must be checked, see arrow
on motor.
If the direction of rotation is incorrect, the grinding will be insuffi-
cient and mechanical components may be damaged.