RKI Instruments NP-204 User Manual

Page 7

Advertising
background image

NP-204

RKI Instruments, Inc.

OPERATION

P r e p a r a t i o n

Normally the NP-204 requires very little preparation before use. Before using the instrument, follow
the steps listed below:

S t a r t - u p

1. Mechanical

Zero

Before turning the instrument on, set it on a flat surface and verify that the meter

reading is exactly on zero. If necessary, turn the Mechanical Zero adjustment to bring

the reading to zero. The mechanical zero adjustment is located centered directly

underneath the meter.

2 .

Battery Check

Turn the range switch to the VOLT ADJ. position. Lift and turn the VOLT ADJ. knob to

bring the meter reading exactly to the red ”Volt Check” line indicated on the meter by

a red “V” and line near the top of the scale. Turning clockwise will increase the voltage

reading.

If you are unable to adjust to the voltage up the red “volt Check” line, replace or

recharge batteries before use. Do not attempt to use the instrument if the

voltage cannot be set high enough.

3 .

Flow System

Attach the hose to the inlet fitting on the left hand side of the instrument, and the

sample probe to the end of the hose. Assure that the hose and probe fittings are secure.

Close off the inlet to probe with finger, and squeeze aspirator bulb. Bulb should remain

collapsed while probe inlet is sealed off with finger. If not, re-check hose and probe

fittings and re-tighten if necessary.

With probe inlet sampling from fresh air, squeeze aspirator bulb 5 times to flush out

any remaining gas from the instrument and sample line.

4.

Zero (Low range)

Turn range selector switch to “L”. Lift and turn the ZERO ADJ. knob to bring the meter

reading to exactly zero. If unable to adjust to zero, filaments may need to be replaced

or sample system may not be fully flushed with fresh air.