Quick Reference Guide For Model Eagle 2

PN 71-0186RK Rev P1

RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587

Phone (510) 441-5656 • (800) 754-5165 • www.rkiinstruments.com

100430/1000

a. Perform a demand zero.

b. Cup hand over end of probe and gently exhale into probe.

c. In a few seconds, the oxygen reading will drop below the alarm

point of 19.5% and activate alarms.

Note: When testing using a hose, allow 1 second of time for each foot of hose

length. For example, allow a 5 foot hose 5 seconds for the sample to reach

the sensors.

d. Press POWER ENTER RESET to reset alarms.

6. Display Mode

a. To access Display Mode, while the Eagle 2 is in normal

operation, press and release DISPLAY ADJUST NO.

b. Use DISPLAY ADJUST NO to scroll through the following

screens: peak readings, battery voltage, gas display, methane

elimination mode (if unit is configured appropriately), relative

response (if RELATIVE RESPONSE is turned on), STEL,

TWA, view alarm settings, select user ID (if turned on), select

station ID (if turned on), time in operation, date/time, and data

logging time remaining (if turned on).

c. Press and release DISPLAY ADJUST NO once more to return

to Normal Operation.

7. Switching from LEL to PPM to %VOL

a. The catalytic combustible channel’s units can be changed from

% LEL (Lower Explosive Limit) to PPM (parts per million) to

% VOL (percent volume).

b. While in normal operation, press and release RANGE ▼

SHIFT until the desired unit appears.

8. Resetting and Silencing Alarms

a. If ALARM SILENCE is turned on, press and release POWER

ENTER RESET to silence an alarm. The audible alarm will

silence and the alarm LED’s will continue to pulse.

b. If the Eagle 2 alarms are set to latching, to reset an alarm,

press and release POWER ENTER RESET once the alarm

condition has cleared.

c. If the Eagle 2 alarms are set to self-resetting, the alarms will

automatically reset once the alarm condition has cleared.

9. Turning Off The Eagle 2

a. Press and hold POWER ENTER RESET. The buzzer will pulse

for about 5 seconds.

b. Release the POWER ENTER RESET button when you see

GOODBYE and the RKI logo. When these disappear, the

Eagle 2 is off.