RKI Instruments GX-82A User Manual
Page 37

Model GX-82A Instruction Manual
34
Hayward, California (510) 441-5656
requirements for portable sample-drawn monitoring.
1. Assemble the probe, hose, bulb, and adapter. Slide the adapter
into place over the sensor cover.
2. Turn on the GX-82A and allow it to complete the start-up proce-
dure.
3. Take the instrument into the area to be monitored.
Caution
Do not place the probe tip in liquid or other materials
which could be drawn into the Aspirator accessory or
the GX-82A.
4. Place the probe tip in the desired location, then squeeze the aspi-
rator bulb several times, at a moderate pace. Watch the GX-82A’s
display for changes.
WARNING
W
ITH
THE
S
AMPLE
-D
RAWING
A
SPIRATOR
ATTACHED
,
THE
GX-82A
DOES
NOT
MONITOR
THE
AMBIENT
AIR
. U
SE
THE
A
SPIRATOR
ONLY
WHEN
SAMPLE
-
DRAWING
MEA-
SUREMENTS
ARE
NECESSARY
,
AND
REMOVE
IT
BEFORE
RESUMING
REGULAR
MONITORING
.