RKI Instruments Beacon 410 User Manual

Page 64

Advertising
background image

60 • Supported Modbus Functions

Beacon 410 Gas Monitor Operator’s Manual

ASCII characters.

Registers 40016 - 40029, Channel 2 Information

Register 40016

Table 15 below shows the information in register 40016.

Registers 40017 & 40018

The 32 bits in registers 40017 and 40018 contain the channel 2 numerical reading in
single precision floating point IEEE format.

Register 40019

The 16 bits in register 40019 contain the channel 2 numerical reading in integer form
omitting the sign (+ or -) and the decimal point.

Register 40020

Table 16 below shows the information in register 40020.

Registers 40021 - 40024

The 8 bytes (characters) in these registers contain the channel gas name in ASCII
characters.

Registers 40025 - 40027

The 6 bytes (characters) in these registers contain the channel gas reading in ASCII
characters.

Table 15: Register 40016

Channel Operational State

(MS-Byte)

Channel Status Bits

(LS-Byte)

•

0 = (this code is not used)

•

1 = Standby

•

2 = Normal Operation

•

3 = Warm Up

•

4 = Fail

•

5 = Post-Calibration

•

6 = Low Power

•

Bit-7 (msb) = Relay-1 (1 = energized)

•

Bit-6 = Relay-2 (1 = energized)

•

Bit-5 = (this bit is not used)

•

Bit-4 = Alarm-1 (1 = asserted)

•

Bit-3 = Alarm-2 (1 = asserted)

•

Bit-2 = Alarm-3 (1 = asserted)

•

Bits[1:0] = Channel Usage Code
[00] = Not Used

[01] = Standby

[10] = Active

[11] = unused code

Table 16: Register 40020

Channel Reading Sign (+/-)

(MS-Byte)

Channel Decimal Point Position

(LS-Byte)

•

0 = Positive (+)

•

1 = Negative (-)

•

0 = No Decimal Point

•

1 = .x

•

2 = .xx

•

3 = .xxx