Sample-Draw Hydrogen Sulfide Gas Detection • 5

Hydrogen Sulfide Transmitter

The hydrogen sulfide transmitter is mounted to the left of the hydrogen sensor. It consists
of the span pot, zero pot, two internally wired terminal strips, and the test points.

Span/zero pots

The span and zero pots are located at the left edge of the transmitter and are used for cali-
bration. Use the span pot to make adjustments to gas response readings and the zero pot
to make adjustments to the zero reading

Termainal Strips

There are two terminal strips which are used for internal factory wiring.

Test points

The test points are located at the top of the transmitter and are labeled

TP+

and

TP-

. A

100 mV - 500 mV output is available at these testpoints for use during calibration.

Main Circuit Board

The main circuit board includes the interconnect terminal strip, transmitter terminal strip,
pump terminal strip, and relay (see Figure 1).

NOTE:

The flowmeter and status lights are mounted to the main circuit board but are

considered part of the flow system.

Interconnect terminal strip

The interconnect terminal strip is the nine-point terminal strip near the bottom edge of the
main circuit board. Use the interconnect terminal strip to connect the sample-draw
detector to power and an external device.

Transmitter terminal strip

The transmitter terminal strip is the nine-point terminal strip near the right edge of the
circuit board. Use the transmitter terminal strip to connect the transmitter to the main
circuit board.

NOTE:

The transmitter is factory-wired to the circuit board. See the “Installation” section

on page 6 for all wiring procedures related to the sample-draw detector.

Pump terminal strip

The pump terminal strip is the four-point terminal strip near the top edge of the circuit
board. Use the pump terminal strip to connect the pump and pressure switch to the main
circuit board.

NOTE:

The pump and pressure switch are factory-wired to the circuit board. See the
“Installation” on page 6 for all wiring procedures related to the sample-draw

detector.

Relay

The relay is to the left of the detector terminal strip. The relay is double pole, double-
throw (DPDT) and is rated for 2 amps at 25 VDC (resistive). If the pressure switch senses a
low flow condition, the relay interrupts the signal from the sensor. The interrupted sensor
signal causes a fail condition at the Pioneer.