Overview
This instruction manual describes the 35-3000RK-OXY sample-draw oxygen detector. This
manual also describes how to install, start up, maintain, and calibrate the sample-draw
detector when using it with a gas monitoring controller. A parts list at the end of this
manual lists replacement parts and accessories for the sample-draw oxygen detector. See
the operator’s manual for the controller for information specific to the controller
Specifications
Table 1 lists specifications for the sample-draw oxygen detector.
Table 1: Specifications
Target Gas
Oxygen (O
2
)
Input Power
24 VDC Nominal (23 VDC - 28 VDC)
Construction (housing)
Fiberglass/polyester (NEMA 4X)
Dimensions
8.5 in. H x 6.5 in. W x 4.25 in. D
Weight
4.5 lbs.
Sampling Method
Sample-draw
Sample Flow
1.5 SCFH (nominal)
Detection Range
0 to 25.0% (by volume)
Response Time
90% in 30 seconds