35-3001A-02 Oxygen Sample-Draw Detector • 19

4.

Verify a voltmeter reading of 100 mV (± 2 mV).

5.

If necessary, use the zero potentiometer on the amplifier to adjust the reading to 100 mV (± 2
mV).

6.

Disconnect the sample tubing from the sample-draw detector’s inlet fitting.

7.

Reconnect the incoming sample line to the inlet fitting.

8.

Unscrew the regulator from the calibration cylinder.

Returning to Normal Operation

1.

Wait approximately one minute to allow the oxygen reading to stabilize.

2.

Remove the voltmeter leads from the amplifier test points.

3.

Close the housing door.

4.

Follow the instructions in the controller’s operator’s manual to exit the calibration mode.

5.

Store the components of the calibration kit in a safe and convenient place.

Parts List

Table 2 lists replacement parts and accessories for the sample-draw detector.

Table 4: Parts List

Part Number

Description

06-1248RK

Sample tubing, 3/16 x 5/16, specify length

06-1248RK-03

Sample tubing, 3/16 in. x 5/16 in., 3 feet (for calibration kit)

07-0053RK

Gasket for oxygen flow block

30-1016RK

Pump

33-0165RK

Hydrophobic filter

33-0167RK

Particle filter

57-1064RK-11

Amplifier

71-0308RK

Operator’s Manual, 35-3001A-02 Oxygen Sample-Draw Detector

81-0076RK-01

Zero air calibration cylinder, 34 liter

81-0076RK-03

Zero air calibration cylinder, 103 liter

81-0078RK-01

Calibration cylinder, 100% nitrogen, 34 liter

81-0078RK-03

Calibration cylinder, 100% nitrogen, 103 liter

81-1054RK

Regulator, demand flow, for 58 and 103 liter steel calibration cylinders

81-1055RK

Regulator, demand flow, for 17 and 34 liter steel calibration cylinders

OS-B11

Oxygen sensor