35-3001A-11 Hydrogen Sample-Draw Detector • 5

counterclockwise to increase the flow and clockwise to decrease the flow.

Status LEDs

Two status LEDs are above the flowmeter. They are also visible through the window in the housing
door. The green Pilot LED is on when the sample-draw detector is receiving power from the
controller. The red Fail LED is on when the sample flow rate is below the low flow level.

Pressure Switch

The pressure switch is mounted to the back of the flowmeter circuit board. The pressure switch
monitors the flow rate of the incoming gas sample.

If the flow rate falls below the preset low flow level, the pressure switch causes the Fail LED to
turn on and interrupts the signal from the detector. The interrupted detector signal causes a fail
condition at the controller. The low flow level is factory-set at 0.6 SCFH (±0.1 SCFH). See
“Adjusting the Low Flow Setting” on page 17 to adjust this setting.

Pump Reset Switch

The pump reset switch is located to the left of the status LEDs. When a low flow condition occurs,
the pump will be shut off. To reset the low flow condition and start the pump again, press and hold
the pump reset switch for about 2 seconds, then release.

Hydrophobic Filter

The hydrophobic filter is located toward the bottom left of the main circuit board. The filter
prevents particulates and water in the incoming gas sample from damaging the flow and detection
systems. Replace the filter when it appears dirty, discolored, or clogged.

Pump

The pump is mounted to the right side of the main circuit board. The pump pulls the gas sample
into the sample-draw detector. The pump operates on 24 VAC, which is generated from the 24
VDC supplied by the controller.

Hydrogen Sensor

The hydrogen sensor detects hydrogen in the %vol range.

Two sensing elements are protected

within the detector. Through a series of thermal and electronic reactions, these elements
produce an electrical output that is proportional to the detection range of the sample-
draw detector.

The hydrogen sensor is installed in the flow block and the flow block is mounted to the middle of
the main circuit board. The sensor housing includes a sintered metal flame arrestor on one end that
allows gas to diffuse into the sensor. On the other end, five pins extend from the sensor and plug
into the socket connector. The connector allows you to replace the sensor without disconnecting the
wiring. Four leads extending from the connector are factory wired to the amplifier.

Amplifier

The amplifier converts the electrical output from the detector to a 4 to 20 mA signal that
corresponds to the detection range and transmits the signal to a gas monitoring controller. A label
on the amplifier indicates the detector drive current. This drive current is factory set and is dictated
by the combustible gas to which the detector is calibrated. Consult RKI Instruments, Inc. if you
plan to change the calibration gas for the sample-draw detector, for example from a hydrogen
calibration to a methane calibration. The amplifier is mounted on the top middle edge of the main
circuit board. It consists of the zero pot, span pot, controller terminal strip, detector terminal strip,
and test points.