Calibration, oxygen detector – RKI Instruments 35-3010RKA-04 User Manual

Page 25

Advertising
background image

35-3010RKA-04 Sample-Draw Detector • 21

5.

Store the components of the calibration kit in a safe and convenient place.

Calibration, Oxygen Detector

This section describes how to calibrate the oxygen detector in the sample-draw detector. It
includes procedures to prepare for calibration, set the zero reading, set the fresh air
reading, and return to normal operation.

NOTE:

This procedure describes calibration using a demand flow regulator, a zero air
calibration cylinder, and a 3-gas calibration cylinder to set the zero reading. A 3-
gas cylinder is used to set the zero since it is already used to calibrate the LEL
channel. A 100% nitrogen calibration cylinder could also be used to set the zero
reading.

Preparing for Calibration

CAUTION:

This procedure may cause alarms at the controller. Take appropriate action to avoid
this, such as entering the calibration mode at the controller.

1.

Open the housing door.

2.

Set a voltmeter to measure in the millivolt (mV) range.

3.

Plug the voltmeter into the test points on the oxygen transmitter. Plug the positive
lead into the test point labeled TP+; plug the negative lead into the test point labeled
TP-

.

4.

Use the following formula to determine the correct test points output for the
calibrating sample.

Output (mV) = (calibrating sample/fullscale) X 400 + 100

For example, with a calibrating sample of 12% and a fullscale setting of 25%, the
correct output for the oxygen test points is 292 mV.

292 (mV) = (12/25) X 400 +100

Setting the Zero Reading

NOTE:

The 3-gas calibration cylinder included in the calibration kit may be used to zero
the oxygen channel. Keep in mind that the oxygen concentration in that cylinder
is 12%, not 0%. Be sure to set the controller appropriately.

1.

Screw the regulator into the 3-gas calibration cylinder.

2.

Connect the calibration kit sample tubing to the regulator.

3.

Allow the sample-draw detector to respond to the calibrating sample for
approximately 1 minute.

4.

After one minute, check the mV output on the oxygen transmitter test points and
verify that the reading matches the response reading (±2 mV) you determined earlier
(292 mV for 12% oxygen).

5.

If necessary, use the span pot on the transmitter to adjust the reading to match the
correct response reading.

6.

Disconnect the sample tubing from the inlet line.

Advertising