Overview, Specifications – RKI Instruments 65-2301RK User Manual
Page 5

65-2300RK/65-2301RK Toxic Detector • 1
Overview
This manual describes the 65-2301RK toxic detector. This manual also describes how to
install, start up, maintain, and calibrate the toxic detector when used with a gas
monitoring controller. A parts list at the end of this manual lists replacement parts and
accessories for the toxic detector.
The 65-2301RK toxic detector includes the 65-2300RK toxic detector and a junction box.
This manual may also be used for the 65-2300RK toxic detector which does not include a
junction box and is normally mounted in one of a controller’s conduit hubs. If you are
using a 65-2300RK toxic detector, disregard all references to the junction box and junction
box terminal strip.
The 65-2300RK and 65-2301RK detectors can be used for various target gases which are
listed in Table 1 below. For a particular target gas, the part number of the toxic detector
will be of the format 65-2300RK-XXX or 65-2301RK-XXX. For example, 65-2301RK-NH3 is
for ammonia (NH
3
) detection.
Specifications
WARNING: Do not use this product in a manner not specified in this instruction
manual.
Table 1 lists specifications for the various 65-2300RK/65-2301RK toxic detectors.
Table 1: Specifications
Target Gas
65-2300RK-ASH3/65-2301RK-ASH3: Arsine (AsH
3
)
65-2300RK-CL2/65-2301RK-CL2: Chlorine (Cl
2
)
65-2300RK-CL-10/65-2301RK-CL-10: Chlorine (Cl
2
)
65-2300RK-HCN/65-2301RK-HCN: Hydrogen Cyanide (HCN)
65-2300RK-NH3/65-2301RK-NH3: Ammonia (NH
3
)
65-2300RK-SO2/65-2301RK-SO2: Sulphur Dioxide (SO
2
)
65-2300RK-PH3/65-2301RK-PH3: Phosphine (PH
3
)
Sampling Method
Diffusion
Detection Range
AsH
3
: 0 - 1.50 ppm
Cl
2
: 0 - 3.00 ppm
Cl
2
: 0 - 10.0 ppm
HCN: 0 - 15.0 ppm
NH
3
: 0 - 75.0 ppm
SO
2
: 0 - 6.00 ppm
PH
3
: 0 - 1.00 ppm
Accuracy
± 10% of reading or ± 5% of full scale (whichever is greater)
Response Time
• Cl
2
and NH
3
: T90 in 90 seconds
• All others: T90 in 45 seconds