Replacing components of the h, S transmitter – RKI Instruments 65-2331RK User Manual

Page 16

Advertising
background image

12 • 65-2331RK H

2

S Transmitter

Table 2: Troubleshooting H

2

S Transmitter

Replacing Components of the H

2

S Transmitter

This section includes a procedure to replace the H

2

S plug-in sensor and amplifier. A

procedure to replace the entire detector assembly is at the end of this section. In most
cases, it is not necessary to replace the entire detector assembly.

Replacing the Plug-In H

2

S Sensor

CAUTION: The sensor contains electrolyte which is a dilute acid. Do not disassemble the sensor

when replacing it with a new one. If sensor electrolyte comes in contact with your
skin, wash affected area thoroughly with soap and water.

1.

Turn off the controller.

2.

Turn off or unplug power to the controller.

Condition

Symptom(s)

Probable Causes

Recommended Action

Fail Condition

• Controller indicates a

fail condition.

• The transmitter wiring

to the controller is
disconnected or
misconnected.

• The plug-in sensor is

not properly plugged
into the sockets in the
detector housing body.

• The wiring from the

detector to the
amplifier is
disconnected or
misconnected.

• The transmitter’s zero

reading is low enough
to cause a fail
condition.

• The transmitter is

malfunctioning.

1. Verify that the transmitter wiring to the

controller is correct and secure.

2. Confirm that the plug-in sensor is

installed properly.

3. Verify that the wiring from the detector

to the amplifier is correct and secure.

4. Perform a fresh air adjustment. A full

calibration is recommended.

5. If the fail condition continues, replace

the plug-in sensor as described later
in this section.

6. If the fail condition continues, contact

RKI for further instruction.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Transmitter responds

slowly or does not
respond to response
test.

• Unable to accurately

set the zero or
response reading
during calibration.

• Transmitter requires

frequent calibration.

Note: Under “normal”
circumstances, the
transmitter requires
calibration once every 3
months.

Some applications

may require a more
frequent calibration
schedule.

• The calibration cylinder

is low, out-dated, or
defective.

• The flame arrestor in

the detector housing
cap is wet or clogged
with dirt or other
particulates.

• The calibration gas is

not an appropriate
concentration.

• The transmitter is

malfunctioning.

1. Verify that the calibration cylinder

contains an adequate supply of a
fresh test sample.

2. Check the detector housing cap to

determine if the flame arrestor is wet
or dirty. Clean if necessary.

3. Verify that the calibration gas

concentration is appropriate for the
transmitter. Zero emission air (20.9%
oxygen) is normally used for a zero
adjustment if the environment is
suspect and 25 PPM H

2

S in nitrogen

is normally used for a response test.

4. If the calibration/response difficulties

continue, replace the plug-in sensor
as described later in this section.

5. If the calibration/response difficulties

continue, contact RKI for further
instruction.

Advertising