Overview, Specifications – RKI Instruments 65-2451RK User Manual

Page 5

Advertising
background image

65-2451RK Hydrogen Transmitter • 1

Overview

This manual describes the 65-2451RK hydrogen transmitter. This manual also describes
how to install, start up, configure, maintain, and calibrate the transmitter when it is used
with a gas monitoring controller. A parts list at the end of this manual lists replacement
parts and accessories for the hydrogen transmitter.

Specifications

Table 1 lists specifications for the hydrogen transmitter.

WARNING: When using the 65-2451RK, you must follow the instructions and warnings

in this manual to assure proper and safe operation of the 65-2451RK and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate the 65-2451RK as described in this manual.

Table 1:Specifications

Description

Specification

Target Gas

Hydrogen

Area Classification

Explosionproof for Class I, Groups B, C, and D

Sampling Method

Diffusion

Signal Output

4 to 20 mA

Detection Range

0 to 100% LEL

Response Time

90% in 45 seconds

Accuracy

± 5% of reading or ± 2% LEL (whichever is greater)