6 • 65-2442RK PPM Hydrogen Transmitter

11. Connect the wires to the applicable detector/transmitter terminal strip at the

controller as shown in Figure 3.

Figure 3: Wiring the PPM Hydrogen Transmitter to a Controller

12. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth)

ground at the controller. RKI controllers typically have a ground stud that can be used
to ground the cable’s drain wire.

Black

Red

Cable Shield

Controller

- (DC Ground)

+ 24 VDC

4 - 20 mA In (S)

Detector W ires