Overview

This manual describes the 65-2498RK hydrogen sulfide (H

2

S) detector. This manual also

describes how to install, start up, maintain, and calibrate the H

2

S detector when used with

a gas monitoring controller. A parts list at the end of this manual lists replacement parts
and accessories for the H

2

S detector.

The 65-2498RK H

2

S detector includes the 65-2495K H

2

S detector and a junction box. This

manual may also be used for the 65-2495RK H

2

S detector which does not include a

junction box and is normally mounted in one of a controller’s conduit hubs. If you are
using a 65-2495RK H

2

S detector, disregard all references to the junction box and junction

box terminal strip.

Specifications

WARNING: Do not use this product in a manner not specified in this instruction

manual.

Table 1 lists specifications for the 65-2495RK and 65-2498RK H

2

S detectors.

Table 1: Specifications

Target Gas

Hydrogen sulfide (H

2

S)

Sampling Method

Diffusion

Detection Range

0 to 100 PPM (parts per million)

Response Time

90% in 30 seconds

Accuracy

± 5% of detection range

Operating Temperature

-4°F to 113°F (-20°C to 45°C)