Calibration frequency – RKI Instruments 65-2504RK User Manual

Page 15

Advertising
background image

65-2504RK Oxygen Transmitter • 11

10. Reconnect the wiring that connects the controller and oxygen transmitter to the

amplifier’s interconnect terminal strip as shown in Table 4 below and Figure 3 on
page 6 of this manual.

11. Reconnect the detector leads to the amplifier’s detector terminal strip as shown in

Table 5 below and Figure 3 on page 6 of this manual.

12. Turn on power to the controller.

13. Turn on the controller.

14. Calibrate the oxygen transmitter as described in “Calibration” on page 12 of this

manual.

Calibration Frequency

Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 3 months is adequate for most oxygen transmitter
applications. Unless experience in a particular application dictates otherwise, RKI
Instruments, Inc. recommends a calibration frequency of every 3 months for the oxygen
transmitter.

If an application is not very demanding, for example detection in a clean, temperature
controlled environment, and calibration adjustments are minimal at calibration, then a
calibration frequency of every 6 months is adequate.

If the application is very demanding, for example if the environment is not well
controlled, then more frequent calibration than every 3 months may be necessary.

Table 4:

Reconnecting the Oxygen Amplifier
to the Controller

Amplifier Interconnect

Terminal Strip

Controller

Transmitter Terminal

Strip

4/20

4 - 20 (FB)

24V

+ (11 VDC - 30VDC)

Table 5:

Reconnecting the Oxygen Detector
to the Amplifier

Detector Lead

Detector Terminal Strip

White

OXY W

Green

OXY G