Troubleshooting – RKI Instruments 65-2515RK User Manual

Page 12

Advertising
background image

8 • 65-2515RK Oxygen Detector

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the oxygen detector.

NOTE:

This troubleshooting guide describes detector problems only. See the controller
operator’s manual for problems you may encounter with the controller.

Replacing the Oxygen Plug-in Sensor

1.

Turn off the controller.

2.

Turn off or unplug incoming power to the controller.

3.

Unscrew the detector housing cap from the detector housing.

4.

Unplug and remove the oxygen sensor.

5.

Carefully plug the replacement sensor into the socket pattern that is located in the top
section of the detector housing.

NOTE:

Match the sensor’s male pins with the two female sockets as you plug the sensor
into the sockets.

6.

Screw the detector housing cap onto the detector housing.

7.

Turn on or plug in incoming power to the controller.

Table 2:Troubleshooting the Oxygen Detector

Condition

Symptom(s)

Probable Causes

Recommended Action

Fail Condition

• Controller indicates a

fail condition.

• The detector wiring is

disconnected or
misconnected.

• The plug-in oxygen

sensor is not properly
installed.

• The detector is

malfunctioning.

1. Verify that the detector wiring is

correct and secure.

2. Confirm that the plug-in oxygen

sensor inside the detector housing is
installed properly.

3. Calibrate the detector.
4. If the fail condition continues, replace

the plug-in oxygen sensor.

5. If the fail condition continues, contact

RKI for further instruction.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Detector responds

slowly or does not
respond to response
test.

• Unable to accurately

set the fresh air or
zero reading during
calibration.

• Detector requires

frequent calibration.

Note: Under “normal”
circumstances, the
detector requires
calibration once every
three months.
Some applications may
require a more frequent
calibration schedule.

• The calibration cylinder

is low, out-dated, or
defective.

• The plug-in sensor is

malfunctioning.

1. Verify that the calibration cylinder

contains an adequate supply of a
fresh test sample.

2. If the calibration/response difficulties

continue, replace the plug-in sensor.

3. If the calibration/response difficulties

continue, contact RKI for further
instruction.

Advertising