43

65-2640RK-03 M2A Transmitter Operator’s Manual

Register 3

Register 3 is the supply voltage (0.1 volt per count).

Registers 4 - 11

The 16 bytes in these registers contain the same ASCII text information that is displayed
on the M2A’s OLED display (not NULL Terminated).

Register 12

Register 12 is the range (full scale readout).

Register 13

Register 13 is the alarm 1 set point.
The decimal point location is the same as specified in Register 1.

Register 14

Register 14 is the alarm 2 set point.
The decimal point location is the same as specified in Register 1.

[8]

Fail Status
0 = Fail Not Asserted
1 = Fail Asserted

[7:6]

Alarm 2 Status
0 = No Alarm
1 = Unacknowledged Alarm
2 = Acknowledged Alarm
3 = Unused Code

[5:4]

Alarm 1 Status
0 = No Alarm
1 = Unacknowledged Alarm
2 = Acknowledged Alarm
3 = Unused Code

[3]

Overscale Status
0 = Not Overscale
1 = Overscale

[2]

Fail Relay Status
0 = Not Energized
1 = Energized

[1]

Alarm 2 Relay Status
0 = Not Energized
1 = Energized

[0]

Alarm 1 Relay Status
0 = Not Energized
1 = Energized

Table 11: Register 2, Operating State, Alarms & Relays Bit & Field Assignments (Continued)

Bit(s)

Value