Start up – RKI Instruments 65-2619RK-CH4-4 User Manual

Page 19

Advertising
background image

65-2619RK-CH4-4 M2 Transmitter Operator’s Manual

14

16. Re-install the control PCB (and ribbon cable if necessary). Be sure the ribbon cable is

routed down below the control PCB so it will not be damaged by the cover when it is
screwed back on.

17. Secure the M2 junction box cover to the junction box.

18. Make controller, device, and relay connections as appropriate. If shielded cable is

used for the PWR/SIG connections, connect the cable shield’s drain wire to an
available chassis ground at the gas monitoring controller, recording device, or
programmable controller.

Start Up

Introducing Incoming Power

1.

Complete the installation procedures described earlier in this manual.

2.

Verify that all wiring connections are correct and secure.

3.

Turn on the incoming power.

4.

If necessary, turn on the controller or other monitoring device that is connected to the
M2.

5.

The LCD display will indicate the firmware version when the M2 is first powered up
and will then count down a one minute warm-up period before normal operation
begins. During normal operation, the display will indicate the target gas and current
gas reading. Verify that the display is indicating the target gas and current gas reading
after the warm-up period is complete and normal operation begins.

NOTE:

Allow the M2’s detector to warm up for 15 minutes before you continue with
the next section, “Adjusting the Fresh Air Reading.”

Adjusting the Fresh Air Reading

When the M2 is shipped from RKI Instruments, Inc., it is factory calibrated. If a full
calibration is desired at startup, see “Calibration” on page 31.

Verify that the M2 is in a fresh air environment (environment known to be free of
combustible or toxic gas vapors and of normal oxygen content, 20.9%).

CAUTION:

If you suspect the presence of combustible gas in the monitoring environment, use
the calibration kit and the zero air calibration cylinder to introduce “fresh air” to
the detector and verify an accurate fresh air setting. See the “Calibration” on
page 31 for instructions on how to use a zero air cylinder when performing a fresh
air adjustment.

Non-Intrusive Fresh Air Signal Adjustment

If the M2 is installed in a classified area and non-intrusive zero adjustment is required,
follow the instructions below, but do not remove the junction box cover. Use the
magnetic wand accessory to actuate the magnetic switches instead of pushing the control
switch buttons. To actuate a magnetic switch and perform the same operation as pressing

METHANE

0 %LEL